Re: [SA_scat] Windows for SAXS cryostat

Dear sir.

I have used Kapton for windows. However, Kapton which I used showed a clear 
scattering peak at a large Q near 0.4 angstrom inverse. If you need the 
scattering data near Q = 0.4 A^-1, you might use UPILEX (Ube polyimide 
file, Ube Industries, ltd., Japan) which shows no scattering peak near Q = 
0.4 A^-1.

Best regards,
T. Kamiyama
_______________________________________________
sa_scat mailing list
sa_scat@iucr.org
http://scripts.iucr.org/mailman/listinfo/sa_scat

Reply to: [list | sender only]
International Union of Crystallography

Scientific Union Member of the International Science Council (admitted 1947). Member of CODATA, the ISC Committee on Data. Partner with UNESCO, the United Nations Educational, Scientific and Cultural Organization in the International Year of Crystallography 2014.

International Science Council Scientific Freedom Policy

The IUCr observes the basic policy of non-discrimination and affirms the right and freedom of scientists to associate in international scientific activity without regard to such factors as ethnic origin, religion, citizenship, language, political stance, gender, sex or age, in accordance with the Statutes of the International Council for Science.